Chapter 9 _______________________________________ Calibration, Adjustment, and Maintenance

2-Point Adjustment

7.

After selecting 2-point adjustment, press

Select to

continue.

8.

Set the probe to a lower reference temperature. You can follow

the stabilization from the graphical display by pressing

Graph. When the reading has stabilized, press

Ready.

9.

Use the arrow buttons to give the lower reference temperature

value and then press

OK.

10.

Set the probe to a higher reference temperature. You can follow

the stabilization from the graphical display by pressing

Graph. When the reading has stabilized, press

Ready.

11.

Use the arrow buttons to give the higher reference temperature

value and then press

OK.

12.

Confirm the adjustment by pressing

Yes. (If you press

No, the adjustment mode display returns and no changes are
made.) The adjustment has been completed.

NOTE

If the difference between the two references is less than
30 ˚C, the adjustment cannot be done.

13.

To exit the adjustment mode, press

Back.

14.

To return to the basic display, press

Exit.