Chapter 9 _______________________________________ Calibration, Adjustment, and Maintenance

17.

Place the screw back onto the calibration button.

18.

To seal the calibration, attach a sticker on the calibration button.

NOTE

Several AutoCals can be necessary after this adjustment has been
performed until the transmitter reaches full accuracy.

1-Point Adjustment

NOTE

This adjustment method allows the customer to manually enter the
reference value when making the adjustment.

9.

A note is displayed on the screen reminding you to follow the

adjustment procedure described in the User's Guide. Press

OK to proceed with the adjustment.

10.

Follow the stabilization from the graphic display by pressing

Graph. The minimum recommended stabilization time for this
critical adjustment is 5 hours.

11.

When the reading has stabilized at the reference level, press

Ready.

12.

Use the arrow buttons to give the reference value. Then press

OK.

13.

Confirm the adjustment by pressing

Yes. If you press

No,

the adjustment mode display returns and no changes are made.

14.

Wait until the automatic auto-calibration has taken place.

15.

When the calibration is completed, press

Back to exit the

adjustment mode.

16.

To return to the basic display, press

Exit.

17.

Place the screw back onto the calibration button.

18.

To seal the calibration, attach a sticker on the calibration button.

NOTE

Several AutoCals can be necessary after this adjustment has been
performed until the transmitter reaches full accuracy.