Chapter 7 ____________________________________________________Sampling from Processes

CHAPTER 7

SAMPLING FROM PROCESSES

When the dewpoint of a process needs to be measured using DM70,
the process can be sampled using one of the following devices:

- DSC74

- DSC74B

- DSC74C

- DSS70A

- DMT242SC

- DMT242SC2

Use the DSC74 sampling cell when the sample is taken from the
pressurized process, where the process pressure forces the sample gas
into the sample cell.

Use the DSC74B sampling cell as DSC74. DSC74B is enhanced
version of DSC74. With version B, harmful gases can be recovered.

DSC74C complements DSC74B with a diffusion coil, which allows
measuring in the atmospheric pressure.

Use the DSS70A sampling system when the process is not
pressurized, that is, the sample must be pumped from the process and
when the process is dirty or too hot for direct measurements.

Use the DMT242SC sampling cell when only the body (thread 3/8"G
and 1/4"G) of the main sampling cell is needed.

Use the DMT242SC2 sampling cell with the welded Swagelok
connectors. This is ideal for sampling in a 1/4" pipeline.

VAISALA _______________________________________________________________________ 47