Table 33, Error messages, Table 33 error messages – Vaisala DMT346 User Manual

Page 147

Advertising
background image

Chapter 6 _______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 147

Table 33

Error Messages

Error Code Error Message

Action

E0

Humidity sensor measurement
malfunction.

Check the integrity of the humidity probe
and the probe cable. Clean the probe
from dirt, water, ice or other
contaminants.

E1

Humidity sensor short circuit

E2

Humidity sensor open circuit

Check the integrity of the humidity probe
and the probe cable.

E3

Temperature sensor open circuit.

E4

Temperature sensor short circuit.

Check the integrity of the humidity probe
and the probe cable. Clean the probe
from dirt water, ice or other
contaminants.

E5

Temperature measurement
malfunction

E6

Temperature sensor current leak.

E7

Internal ADC read error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E9

Checksum error in the internal
configuration memory

E10

Internal EEPROM read error

E11

Internal EEPROM write error

E12 … E13 Add-on module 1 (or 2) connection

failure

Turn off the power and check the
module connection. Turn on the power.

E14

Device internal temperature out of
range

Ensure that the operating temperature is
within the valid range.

E15

Internal RAM memory or software
checksum error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E18

Internal ADC reference voltage out
of range

E19

Internal analog output reference
voltage out of range

E20 … 22

Configuration switches for analog
output 1/2/3 set incorrectly

Check and re-set the switches, see
page 57.

E24 … E25 Internal error in add-on module 1 (or

2)

Disconnect the power and check the
module connection.

E26

Communication module installed in
incorrect add-on module slot

Disconnect the power and change the
communication module to another
module slot.

E28 … E29 Unknown/incompatible module

installed in add-on module slot
1 (or 2)

Ensure that the module is compatible
with the DMT345/346.

E30

Internal analog voltage out of range Internal transmitter failure. Remove the

transmitter and return the faulty unit to
Vaisala Service.

E31

Internal system voltage out of range Check that the supply voltage is correct

and the power supply can provide
enough power for the instrument.