Vaisala DMT346 User Manual

Page 154

Advertising
background image

User's Guide _______________________________________________________________________

154 __________________________________________________________________ M210762EN-G

11. When stabilized, type the high end reference value after the

question mark and press Enter.

>fcrh

RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...

RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>

12. The OK indicates that the adjustment has succeeded and the new

calibration coefficients are calculated and stored. Enter the
adjustment information (date and text) to the memory of the
transmitter, see section Adjustment Information on page 151.

13. Press the ADJ button on the motherboard to close the adjustment

mode.

14. Take the probe out of the reference conditions and replace the

filter.