List of Serial Commands

All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.

The notation <cr> refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.

In the tables below, the bold text in the brackets is the default setting.

Table 18

Measurement Commands

Command

Description

R

Start the continuous outputting

S

Stop the continuous outputting

INTV [0 ... 255 S/MIN/H]

Set the continuous output interval
(for RUN mode)

SEND [0 ... 255]

Output the reading once

DSEND

Output the reading once from all connected
transmitters (in STOP or POLL mode)

SMODE
[STOP/RUN/POLL/MODBUS]

Set the serial interface mode

SERI [baud p d s]

User Port settings (Default: 4800 E 7 1)

ADDR [0 ... 255]

Set the transmitter address (for POLL and
MODBUS modes)

NET

View or set networking parameters for LAN
and WLAN interfaces

WLAN

View or set wireless network parameters
for WLAN interface

OPEN [0 ... 255]

Open temporarily connection to a POLL
mode device

CLOSE

Close the temporary connection (back to
POLL Mode)

Table 19

Formatting Commands

Command

Description

FORM

Set the output format of SEND and R
commands

FST

Add the state of AutoCal, purge and sensor
warming in connection with SEND and R
commands

SCOM

Assign a new command name that works
like the SEND command

FDATE

Add date to R and SEND outputs

FTIME

Add time to output to SEND and R outputs