O figure 22 on pa, Figure 23 on, E 38) – Vaisala DMT346 User Manual

Page 38: Figure 23 installing probe through thick walls

Advertising
background image

User's Guide _______________________________________________________________________

38 ___________________________________________________________________ M210762EN-G

5.

Lock the probe in place by tightening the locking screws on the
bar.

0604-024

Figure 22

Installing the Probe through Process Wall

0604-025

Figure 23

Installing Probe through Thick Walls

120

cooling bar

sealing material
(for example,
mineral wool)

PROCESS
(t

max

= 350 °C)

mounting flange

to be
welded

cooling
sleeve

ø

89.

5 +

0.

5

DRYCAP

®

sensor

O-ring FPM

protective
tube

channel with
mineral wool

125

(max. 175 mm)

PROBE

cooling
fins

locking screws
of the probe
M4×6 DIN 916
(AISI 316)

marking
hole

8

40.6

to be welded

ø

89.

5 +

0.

5

lengthening
piece

inner plate of the
process wall

wall thickness

< 160

plate ring

plate pipe

the flange is fixed
with screws or by
spot welding

cooling
profile

~45º

ø

118

NOTE:
Make sure that
the cooling profile
does not cover the
locking screws