Table 36, Temperature specifications dmt345, Table 37 – Vaisala DMT346 User Manual

Page 162: Relative humidity specifications dmt345, Table 38, Mixing ratio specifications dmt345, Table 39, Dewpoint specifications dmt346

Advertising
background image

User's Guide _______________________________________________________________________

162 __________________________________________________________________ M210762EN-G

Table 36

Temperature Specifications DMT345

Property

Value

Measurement range

0 ... +180 °C (+32 ... +356 °F)

with sensor warming

upper range limited by humidity (at
80 %RH warming is switched on and
T reading not actual process
temperature)

Accuracy

±0.4 °C at 100 °C

Temperature sensor

Pt 100 IEC 751 1/3 class B

Table 37

Relative Humidity Specifications DMT345

Property

Value

Measurement range

0 ... 100 %RH

with sensor warming

0 ... 80 %RH

Accuracy

below 10 %RH

±10 % of reading

above 10 %RH

±1.5 %RH + 1.5 % of reading

Table 38

Mixing Ratio Specifications DMT345

Property

Value

Measurement range (typical)

0 ... 1000 g/kg (0..7000 gr/lbs)

Accuracy

±12 % of reading

Table 39

Dewpoint Specifications DMT346

Property

Value

Sensor

Vaisala DRYCAP®180S

Measurement range

-25 … +100 °C (-13 … +212 °F) T

d

Accuracy

±2 °C (±3.6 °F) T

d

See the accuracy graph in Figure 79
on page 163

Response time 63% [90%] flow rate
1l/min and 1 bar pressure

from dry to wet

5s [10 s]

from wet to dry including auto-
calibration

45s [5 min]

Advertising