Operating environment – Vaisala PTB330 User Manual

Page 128

Advertising
background image

USER'S GUIDE____________________________________________________________________

126 _________________________________________________________________ M210855EN-D

Table 22

Temperature Dependence****

Class B

500 ... 1100 hPa

±0.1 hPa

50 ... 1100 hPa

±0.3 hPa

Table 23

Total Accuracy at -40 ... +60°C (-40 ... +140 °F)

Class A

Class B

500 ... 1100 hPa

±0.15 hPa

±0.25 hPa

50 ... 1100 hPa

±0.45 hPa

Table 24

Long Term Stability

500 ... 1100 hPa

±0.10 hPa/year

50 ... 1100 hPa

±0.20 hPa/year

*Defined as ±2 standard deviation limits of endpoint non-linearity,
hysteresis or repeatability error.

**Defined as ±2 standard deviation limits of inaccuracy of the
working standard including traceability to NIST.***Defined as the
root sum of the squares (RSS) of endpoint non-linearity, hysteresis
error, repeatability error, and calibration uncertainty at room
temperature ****Defined as ±2 standard deviation limits of
temperature dependence over the operating temperature range.

Operating Environment

Table 25

Operating Environment

Pressure range

500 ... 1100 hPa, 50 ... 1100 hPa

Temperature range

Operating without display

-40 ... +60°C (-40 ... +140 °F)

Operating with display

0 ... +60°C (-32 ... +140 °F)

Storage without display

-55 ... +80 °C (-67 ... +176 °F)

Storage with display

-40 ... +80 °C (-40 ... +176 °F)

Maximum pressure limit

5000 hPa abs.

Electromagnetic compatibility EMC standard EN61326-1:1997+ Am1:1998

+ Am2:2001: Industrial Environment