Vers, Resetting barometer by using serial line, Reset – Vaisala PTB330 User Manual

Page 90

Advertising
background image

USER'S GUIDE____________________________________________________________________

88 __________________________________________________________________ M210855EN-D

NOTE

Because ERRS command does not display any difference between
present and past errors, execute ERRS twice in a row to see which
ones are still present (past errors are not displayed anymore on the
second ERRS response).

Table 17

ERRS command line parameters

Value Description

Display currently active and past, unviewed (unacknowledged)
errors.

LOG

Show error log contents.

BARO

Show detailed error information for pressure transducer errors.


Example:

>errs
PASS
No errors
>errs
FAIL
Error: Operating voltage out of range.
>

VERS

Use the command VERS to display software version of PTB330:

>VERS
MT300 / 1.13
>

Resetting Barometer by Using
Serial Line

RESET

Resets the device. The user port switches to start-up output mode
selected with command SMODE.