Vaisala PTB330 User Manual

Page 71

Advertising
background image

Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 69

Measurement setting commands
TQFE

[-80...+200° C]

Shows or sets the temperature for QFE
corrected pressure.

DPMAX

[0…9999.99 hPa] Shows or sets the maximum permissible

difference pressure between barometer
modules.

HHCP

[ -30...+30 m]

Shows or sets the altitude for height
corrected pressure.

HQFE

[ -30...+30 m]

Shows or sets the altitude for the QFE
corrected pressure.

HQNH

[ -30...+3000 m]

Shows or sets the altitude for the QNH
corrected pressure.

PSTAB

[0…9999.99 hPa] Shows or sets the pressure stability limits.

AVRG

[0…600 s]

Sets the barometer measurement
averaging time (in seconds).

Formatting commands
FORM

Sets the custom output for the SEND
command and for RUN mode.

TIME

Shows or changes the current time.

DATE

Shows

or

changes

the current date.

UNIT

Shows or sets unit for a quantities.

Data recording commands
DSEL

Selects the quantities that are displayed on
the graphical user interface as well as
quantities for data recording.

DELETE

Erases the log memory.

UNDELETE

Restores the erased log memory.

DIR

Lists the available logs in the logging
memory.

PLAY

Shows the trend, min, and max values of
the given log.

Calibration and adjustment commands
CDATE

Shows or sets the calibration date.

LCP1/
LCP2/
LCP3

Performs a linear adjustment for the
barometer module/module.

MPCP1/M
PCP2/MP
CP3

Performs a multipoint adjustment for the
barometer module/module.

CTEXT

Shows or sets the calibration info text.

NOTE

The calibration and adjustment commands are available in adjustment
mode only. Press the adjustment button before inputting these
commands.