Vaisala PTB330 User Manual

Page 76

Advertising
background image

USER'S GUIDE____________________________________________________________________

74 __________________________________________________________________ M210855EN-D

UNIT [x] [y]

where:

x = Output

quantity

y = Output

unit

Use the UNIT command to change the measurement unit for the
quantities. The output consists of all the measured quantities of the
unit you enter as variable [y].

Example:

>unit Pa
P : Pa
P3h : Pa
P1 : Pa
P2 : Pa
DP12 : Pa
HCP : Pa
QFE : Pa
QNH : Pa
>

Use the UNIT command with a quantity symbol to change the
measurement unit for a specific quantity. Specify the quantity and the
desired unit for it.

Example:

>unit P mmhg
P : mmHg
P3h : Pa
P1 : Pa
P2 : Pa
DP12 : Pa
HCP : Pa
QFE : Pa
QNH : Pa
>

NOTE

This command changes both the serial output and display units.

Use the UNIT ?? command to list the available measurement units for
the quantities.

Advertising