Measuring settings, Diagnostic settings – Vaisala PTB330 User Manual

Page 58

Advertising
background image

USER'S GUIDE____________________________________________________________________

56 __________________________________________________________________ M210855EN-D

Measuring Settings

See the calculation formulas for pressure values in Appendix A,
Calculation Formulas, on page 133. To change the m
easuring settings:

1. Open

the

MAIN MENU by pressing any of the arrow buttons.

2. Select

Measuring and press the right arrow button.

3. Select

Measuring settings. Press the right arrow button.

4.

Select an option with up/down arrow buttons. Press the SET
button.

5.

First press the left/right arrow buttons to move from a digit to
another. Then press the up/down arrow buttons to enter desired
numeric value for the digit. Repeat the procedure until all the
desired digits are set. Press the OK button.

6. Press

EXIT to return to the basic display.

Diagnostic Settings

Use the display/keypad or the serial line commands PSTAB
on page 82, and
DPMAX on page 78 to set the diagnostic settings:

1.

Press any of the arrow buttons to open the MAIN MENU.

2. Select

System and press the right arrow button to confirm your

selection.

3. Select

Diagnostics and press the right arrow button to confirm

your selection.

4. Select

Diagnostic Settings.

5.

Use the up/down arrow buttons to select Stability. Press SET
to input the value. Use the left/right arrow buttons to move
between the value and the unit. Set the value and the unit using
the up/down arrow buttons. Press OK to confirm your choice.

6.

Proceed as instructed in point 5 to set Max. diff.

7. Press

SET to confirm your selection.

8. Press

EXIT to return to the basic display.

Use the display/keypad to check the stability of the pressure
measurement. If concurrent pressure measurements differ less than the
set value, the display outputs a small "OK" text.