Setting communication hardware – Vaisala GMP343 User Manual
Page 35

Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 33
In STOP mode: Measurements output only by command, all commands
can be used (default mode).
In RUN mode: Outputting automatically, only command S can be used.
In POLL mode: Measurements output only with command SEND. For
more details on POLL mode, see section
Networking Operation on page 37
.
The settings are not valid until the device is reset.
Setting Communication Hardware
Communication hardware can be chosen between RS-232 or RS-485.
Command parameters are correspondingly 232 and 485.
RSMODE <cr>
SAVE <cr>
The setting is not valid until the device is reset.
>rs mode
RSMODE : 232? 485
>save
EEPROM saved successfully.
>