Relative humidity compensation on/off – Vaisala GMP343 User Manual

Page 53

Advertising
background image

Chapter 4 ________________________________________________________________ Operation

VAISALA_______________________________________________________________________ 51

Relative Humidity Compensation ON/OFF

To enable or disable the humidity compensation, use the commands:

RHC x <cr>

SAVE <cr>

x=ON/OFF (default =ON)

>rhc on

RHC : ON

>rhc off

RHC : OFF

>

>save

EEPROM saved successfully.

>

NOTE

Check that the ambient pressure value is correct. The correct pressure
value is needed for RH compensation even if the pressure compensation
is disabled.

Advertising