Temperature Compensation ON/OFF
The internal temperature sensor is located in the measurement chamber.
Temperature compensation is done automatically unless for some reason
you want to take the compensation feature off.
To enable or disable temperature compensation, use the commands:
TC x <cr>
SAVE <cr>
x=ON/OFF (default =ON)
>tc on
TC : ON
>tc off
TC : OFF
>
>save
EEPROM saved successfully.
>
Setting the Oxygen Concentration
To set the oxygen concentration value, use the commands:
O x <cr>
SAVE <cr>
x=0...100 % (default =20.95 %)
>o 21
OXYGEN (%): 21.00
>save
EEPROM saved successfully.
>