Setting the analog output error level – Vaisala GMP343 User Manual
Page 58
Advertising

User’s Guide ______________________________________________________________________
56 ___________________________________________________________________M210514EN-E
Setting the Analog Output Error Level
Factory default states for analog outputs during error condition are the
maximum values of the measuring range=2.5 V / 5 V / 20 mA.
AERR x <cr>
SAVE <cr>
x=error value (mA) or (V)
Acceptable ranges: 0...20 mA, 0...6 V
Example:
>aerr
AERR (V) : 5.000 ? 0
>
>save
EEPROM saved successfully.
>
NOTE
The error output value must be within a valid range for current (0...20
mA) or voltage output (0...6 V) regardless of the UHIGH setting.