Setting the high limit of the concentration range, Setting the low limit of the concentration range – Vaisala GMP343 User Manual
Page 56
Advertising

User’s Guide ______________________________________________________________________
54 ___________________________________________________________________M210514EN-E
Setting the High Limit of the
Concentration Range
Sets the hight CO
2
concentration value, which corresponds to the analog
output high value.
AHIGH <cr>
SAVE <cr>
Key in the highest concentration value (ppm) and press ENTER.
>ahigh
AHIGH (ppm) : 1000.0 ? 1200
>
>save
EEPROM saved successfully.
>
Setting the Low Limit of the
Concentration Range
Sets the low CO
2
concentration value, which corresponds to the analog
output low value.
ALOW <cr>
SAVE <cr>
Key in the lowest concentration value (ppm) and press ENTER.
>alow
ALOW (ppm) : 0.0 ? 20
>
>save
EEPROM saved successfully.
>