Chapter 4 ____________________________________________________ Vaisala Industrial Protocol

VAISALA ________________________________________________________________________ 29

Table 8

Advanced Serial Commands

Command

Description

ADDR [0 … 254]

Show or set probe address.

AMODE

Show or set analog output mode.

AOVER

Show or set analog output overrange and
clipping behavior.

ASEL

Show or set analog output parameter.

ATEST

Test analog putput.

CCO2

Adjust CO

2

measurement gain and offset.

CDATE

Show or set calibration date.

CP

Adjust pressure measurement offset.

CT

Adjust temperature measurement offset.

CTEXT

Show or set calibration information.

FRESTORE

Restore probe to factory settings.

O2CMODE

Show or set oxygen compensation mode.

PCMODE

Show or set pressure compensation mode.

RHCMODE

Show or set humidity compensation mode.

TCMODE

Show or set temperature compensation mode.

Showing Device Information

Show Probe Information

The ? command outputs a listing of probe information.

?<cr>

Command ?? is a similar command that is designed to be used if the
probe is in POLL mode but its address is unknown. The probe will
respond to the ?? command even while in POLL mode. Do not use the ??
command if you have more than one probe on an RS-485 line.

Example:

?
Device : GMP231
Copyright : Copyright (c) Vaisala Oyj 2013. All rights reserved.
SW Name : GMP231
SW version : 1.0.1.1537
Snum : J1320082
Calibrated : 2014-03-11
Address : 1
Smode : STOP