Example: set compensation temperature, Table 28, Example set_parameter invoke message – Vaisala GMP231 User Manual
Page 64: Table 29, Example set_parameter response message
Advertising

User's Guide _______________________________________________________________________
62 ___________________________________________________________________ M211501EN-C
Example: Set Compensation
Temperature
Table 28
Example Set_Parameter Invoke Message
Message Bytes (hex) Content
12
I
2
C address
82
Set_parameter command
09
Device address
0A
Frame length
10
Parameter ID of compensation temperature
00
Temperature 37C
00
14
42
0C
CRC high
6E
CRC lo
Table 29
Example Set_Parameter Response Message
Message Bytes (hex) Content
13
I
2
C address
04
Status byte (error active, ACK)
82
Set_parameter command
09
Device address
08
Frame length
10
Parameter ID of compensation temperature
00
Return code: Ok
D1
CRC high
CE
CRC lo
Advertising