Changing the silicone plug, Figure 19 – Vaisala GMP231 User Manual

Page 74

Advertising
background image

User's Guide _______________________________________________________________________

72 ___________________________________________________________________ M211501EN-C

Changing the Silicone Plug

Change the silicone plug if it shows visible signs of contamination. It is
best to replace the filter also.

1.

Pull the white PTFE filter away from the sensor. Rotating the filter
slightly makes it easier to detach

2.

Lift one edge of the plug and pull it away from the installation tube.

3.

Use clean instrument air to gently blow any loose dirt and filter
material away from the sensor.

4.

Clean the installation tube and the wall around the plug.

5.

Place the new silicone plug over the sensor and push it into the
installation tube.

6.

Take a new clean filter and push it onto the sensor. Push and rotate
the filter slightly to make sure it is fully in place.

1403-157

Figure 19

Changing the Silicone Plug