Agilent Technologies 34401A User Manual

Page 128

Advertising
background image

<

function

>:RESolution {<

resolution

>|MINimum|MAXimum}

Select the resolution for the specified function (not valid for frequency,
period, or ratio). Specify the resolution in the same units as the
measurement function, not in number of digits.

MIN

selects the smallest

value accepted, which gives the most resolution.

MAX

selects the largest

value accepted which gives the least resolution. [Stored in volatile memory]

<

function

>:RESolution? [MINimum|MAXimum]

Query the resolution for the selected function. For frequency or period
measurements, the multimeter returns a resolution setting based upon
a 3 Hz input frequency.

<

function

>:NPLCycles {0.02|0.2|1|10|100|MINimum|MAXimum}

Select the integration time in number of power line cycles for the
present function (the default is 10

PLC

). This command is valid only for

dc volts, ratio, dc current, 2-wire ohms, and 4-wire ohms.

MIN

= 0.02.

MAX

= 100. [Stored in volatile memory]

<

function

>:NPLCycles? [MINimum|MAXimum]

Query the integration time for the selected function.

FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}

Select the aperture time (or gate time) for frequency measurements
(the default is 0.1 seconds). Specify 10 ms (4

1

2

digits), 100 ms (default;

5

1

2

digits), or 1 second (6

1

2

digits).

MIN

= 0.01 seconds.

MAX

= 1 second.

[Stored in volatile memory]

FREQuency:APERture? [MINimum|MAXimum]

Query the aperture time for frequency measurements.

PERiod:APERture {0.01|0.1|1|MINimum|MAXimum}

Select the aperture time (or gate time) for period measurements
(the default is 0.1 seconds). Specify 10 ms (4

1

2

digits), 100 ms (default;

5

1

2

digits), or 1 second (6

1

2

digits).

MIN

= 0.01 seconds.

MAX

= 1 second.

[Stored in volatile memory]

PERiod:APERture? [MINimum|MAXimum]

Query the aperture time for period measurements.

Chapter 4 Remote Interface Reference
Measurement Configuration Commands

122