To Make Null (Relative) Measurements

Each null measurement, also called relative, is the difference between a
stored null value and the input signal.

Result = reading – null value

•

You can make null measurements with any function except
continuity, diode, or ratio. The null operation is local to the selected
function; when you change functions, null is disabled.

•

To null the test lead resistance for more accurate two-wire ohms
measurements, short the ends of the test leads together and then
press

Null

.

•

The first reading taken after you press

Null

is stored as the null

value in the Null Register. Any previously stored value is
replaced with the new value.

•

After enabling null, you can edit the stored null value by
pressing

Shift >

(Menu Recall). This takes you to the

“

NULL VALUE

” command in the

MATH MENU

(only if null is

enabled). Go down to the “parameter” level, and then edit the
displayed value.

•

The null register is cleared when you change functions, turn null off,
turn off the power, or perform a remote interface reset.

Enables null operation;
Press again to disable.

Math annunciator is on when
null operation is enabled.

To read / edit the null value, use the MATH menu.

Chapter 2 Front-Panel Menu Operation
To Make Null (Relative) Measurements