Baud Rate Selection (

RS-232

)

You can select one of six baud rates for

RS-232

operation. The rate is set

to 9600 baud when the multimeter is shipped from the factory.

The baud rate can be set only from the front-panel.

•

Select one of the following: 300, 600, 1200, 2400, 4800, or 9600 baud
(factory setting).

•

The baud rate selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

•

Front-Panel Operation:

3: BAUD RATE (I/O MENU)

See also “To Set the Baud Rate,” on page 163.

Parity Selection (

RS-232

)

You can select the parity for

RS-232

operation. The multimeter is

configured for even parity with 7 data bits when shipped from the factory.

The parity can be set only from the front-panel.

•

Select one of the following: None (8 data bits), Even (7 data bits), or
Odd (7 data bits). When you set the parity, you are indirectly setting
the number of data bits.

•

The parity selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

•

Front-Panel Operation:

4: PARITY (I/O MENU)

See also “To Set the Parity,” on page 164.

3

Chapter 3 Features and Functions
Remote Interface Configuration

93