Baud Rate Selection (

RS-232

)

You can select one of six baud rates for

RS-232

operation. The rate is set

to 9600 baud when the multimeter is shipped from the factory.

The baud rate can be set only from the front-panel.

Select one of the following: 300, 600, 1200, 2400, 4800, or 9600 baud
(factory setting).

The baud rate selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

Front-Panel Operation:

3: BAUD RATE (I/O MENU)

See also “To Set the Baud Rate,” on page 163.

Parity Selection (

RS-232

)

You can select the parity for

RS-232

operation. The multimeter is

configured for even parity with 7 data bits when shipped from the factory.

The parity can be set only from the front-panel.

Select one of the following: None (8 data bits), Even (7 data bits), or
Odd (7 data bits). When you set the parity, you are indirectly setting
the number of data bits.

The parity selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

Front-Panel Operation:

4: PARITY (I/O MENU)

See also “To Set the Parity,” on page 164.