Integration Time

(continued)

•

Front-Panel Operation: Integration time is set indirectly when you
select the number of digits. See the table for resolution on page 54.

•

Remote Interface Operation:

<

function

>:NPLCycles {0.02|0.2|1|10|100|MINimum|MAXimum}

For frequency and period measurements, aperture time (or gate time)
is analogous to integration time. Specify 10 ms (4

1

⁄

2

digits), 100 ms

(default; 5

1

⁄

2

digits), or 1 second (6

1

⁄

2

digits).

FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
PERiod:APERture {0.01|0.1|1|MINimum|MAXimum}

Front / Rear Input Terminal Switching

Any measurement made using the front terminals can also be made
using the input terminals on the rear panel. See “The Front Panel at
a Glance,” on page 2, for the location of the front / rear switch.

The input terminals can only be configured from the front panel.
You cannot select the terminals from the remote interface, but you
can query the present setting.

•

The

Rear

annunciator turns on when you select the rear terminals.

•

Remote Interface Operation: You can query the multimeter to
determine whether the front or rear input terminals are selected.

ROUTe:TERMinals?

returns “

FRON

” or “

REAR

”

Chapter 3 Features and Functions
Measurement Configuration

58