Agilent Technologies 34401A User Manual

Page 75

Advertising
background image

Limit Testing

The limit test operation enables you to perform pass/fail testing to
upper and lower limits that you specify.

Applies to all measurement functions, except continuity and diode tests.

•

You can set the upper and lower limits to any value between 0 and

±

120% of the highest range, for the present function. The upper limit

selected should always be a more positive number than the lower
limit. The default upper and lower limits are both “0”.

•

The upper and lower limits are stored in volatile memory; the
multimeter sets both limits to “0” when power has been off, after a
remote interface reset, or after a function change.

•

You can configure the multimeter to generate a service request (

SRQ

)

on the first occurrence of a failed reading. See “The SCPI Status
Model,” starting on page 134 for more information.

•

Front-Panel Operation: The multimeter displays “

OK

” for each

reading that is within the specified limits. It displays “

HI

” or “

LO

” for

each reading that exceeds the upper or lower limit. The multimeter
beeps once (if the front-panel beeper is enabled) on the first
occurrence of a failed reading after a good reading. See also “Beeper
Control,” on page 88.

5: LIMIT TEST (MATH MENU)

enable or disable limit test

6: HIGH LIMIT (MATH MENU)

set the upper limit

7: LOW LIMIT (MATH MENU)

set the lower limit

You can also turn off limit test by selecting a different math operation
from the front panel (only one math operation can be enabled at a time).

3

Chapter 3 Features and Functions
Math Operations

69

Advertising