Automatic trigger delays – Agilent Technologies 34401A User Manual

Page 87

Advertising
background image

Automatic Trigger Delays

If you do not specify a trigger delay, the multimeter selects an
automatic delay for you. The delay is determined by function, range,
integration time, and ac filter setting.

•

DC Voltage and DC Current (for all ranges):

•

Resistance (2-wire and 4-wire):

•

AC Voltage and AC Current (for all ranges):

•

Frequency and Period:

Integration Time

NPLC

≥

1

NPLC

<

1

Trigger Delay

1.5

ms

1.0

ms

Range

100

Ω

1 k

Ω

10 k

Ω

100 k

Ω

1 M

Ω

10 M

Ω

100 M

Ω

Trigger Delay
(For NPLC

≥

1)

1.5 ms

1.5

ms

1.5

ms

1.5

ms

15

ms

100

ms

100

ms

AC Filter

Slow
Medium
Fast

Trigger Delay

7.0

sec

1.0

sec

600

ms

Remote or single/external trigger

Range

100

Ω

1 k

Ω

10 k

Ω

100 k

Ω

1 M

Ω

10 M

Ω

100 M

Ω

Trigger Delay
(For NPLC < 1)

1.0

ms

1.0

ms

1.0

ms

1.0

ms

10

ms

100

ms

100

ms

AC Filter

Slow
Medium
Fast

Trigger Delay

1.5

sec

200

ms

100

ms

Front panel with auto trigger ON

Trigger Delay

1.0 sec

Remote or single/external trigger


Trigger Delay

0 sec

Front panel with auto trigger ON

3

Chapter 3 Features and Functions
Triggering

81

Advertising