Appendix G. CPEC200 Scrub Module Installation, Operation and Maintenance

Disconnecting this tube ensures the bottles are not

pressurized when the cover is removed. The scrub module

has been designed to require this tube to be disconnected

before removing the cover as a safety precaution.

4. Loosen the four thumbscrews (shown in FIGURE G-2) and remove

the cover plate to gain access to the bottles (FIGURE G-3). Note that

the thumb screws are captive; they remain attached to the cover plate.

FIGURE G-3. Interior of CPEC200 scrub module with tubing and cover

removed

5. Disconnect the remaining tubes from the bottles at the Swagelok

®

fittings.

Caps are spring loaded!

6. Remove the center bottle from the scrub module and set it aside.
7. Remove the left bottle and place it in the center position.
8. Remove the right bottle and place it in the left position.
9. Refill the bottle that was removed by twisting while pulling to remove

the top cap. The caps are held in place by friction only and the spring

inside the bottle may eject the cap.

NOTE

NOTE

G-4