Campbell Scientific CPEC200 Closed-Path Eddy-Covariance System User Manual
Page 102
Appendix G. CPEC200 Scrub Module Installation, Operation and Maintenance
Disconnecting this tube ensures the bottles are not
pressurized when the cover is removed. The scrub module
has been designed to require this tube to be disconnected
before removing the cover as a safety precaution.
4. Loosen the four thumbscrews (shown in FIGURE G-2) and remove
the cover plate to gain access to the bottles (FIGURE G-3). Note that
the thumb screws are captive; they remain attached to the cover plate.
FIGURE G-3. Interior of CPEC200 scrub module with tubing and cover
removed
5. Disconnect the remaining tubes from the bottles at the Swagelok
®
fittings.
Caps are spring loaded!
6. Remove the center bottle from the scrub module and set it aside.
7. Remove the left bottle and place it in the center position.
8. Remove the right bottle and place it in the left position.
9. Refill the bottle that was removed by twisting while pulling to remove
the top cap. The caps are held in place by friction only and the spring
inside the bottle may eject the cap.
NOTE
NOTE
G-4