Appendix A. CPEC200 Diagnostics

because the EC100 is not powered, or the SDM cable from the

datalogger to the EC100 is not connected.

ERROR: No EC155 detected - Check EC155 connections to EC100 is

displayed if

diag_irga = −1 (all diagnostic bits are set). This usually

means the EC155 sensor head is not connected to the EC100.

If bit 9 of

diag_irga is set, this indicates the EC155 has been powered

down. There are multiple possible causes and each has a different

message reported in

irga_status:

o ERROR: IRGA is OFF - Check BattVolt is displayed if

BattVolt_OK = False. This usually means the EC155 has

been powered down by the CPEC200 because the battery

voltage has dropped below the shutdown limit.

o ERROR: IRGA is OFF - Check source_tmpr is displayed if

BattVolt_OK = True and bit 7 of diag_irga is set. This

usually indicates the EC155 has powered itself down because

the internal temperature,

source_tmpr, is outside its

operating range. The EC155 will power itself off if

source_tmpr rises above 55°C, and will not turn back on

until the temperature drops to 50°C. At cold temperatures,

the EC155 will power off at –35°C and back on at –30°C.

o ERROR: IRGA is OFF is displayed if the EC155 is powered

down for an unknown reason.

IRGA is starting up is displayed if bit 3 of

diag_irga is set. This

normally indicates the EC100 has just been powered up or the EC155

has been turned on. It may take a few minutes for the IRGA to start

up and report data.

ERROR: diff_press problem - Check intake filter is displayed if bit 23

of

diag_irga is set. Normally, this indicates the EC155 intake filter

has plugged, which causes a pressure drop in the EC155 greater than

the differential pressure sensor’s ± 7 kPa range.

ERROR: CO2 signal is too low - Check EC155 windows is displayed

when the CO

2

signal level is below 0.8. This normally indicates the

EC155 windows have become dirty and should be cleaned.

ERROR: H2O signal is too low - Check EC155 windows is displayed

when the H

2

O signal level is below 0.8. As with CO

2

signal level, this

normally indicates the EC155 windows have become dirty and should

be cleaned.

ERROR: IRGA problem - Check diag_irga is displayed when some

other bit is set in

diag_irga. See the EC155 user manual for further

troubleshooting suggestions.

sonic_status

Public variable

sonic_status gives the status of the sonic anemometer and is

based on the diagnostic word from the CSAT3A (

diag_sonic).

Sonic is OK is displayed if

diag_sonic = 0. This means the CSAT3A

has detected no problems.

ERROR: No data from sonic - Check EC100 power and

communication is displayed if

diag_sonic = NAN. This indicates the

EC100 does not respond to the datalogger’s request for data, usually

because the EC100 is not powered, or the SDM cable from the

datalogger to the EC100 is not connected.

ERROR: No CSAT3A detected - Check CSAT3A connections to

EC100 is displayed if

diag_sonic = −1 (all diagnostic bits are set).