8. wiring of power and communications, 1. sdm wiring – Campbell Scientific CPEC200 Closed-Path Eddy-Covariance System User Manual

Page 35

Advertising
background image

CPEC200 Closed-Path Eddy-Covariance System

Wire the SDM communications cable (CABLE4CBL-L) between the EC100

and the CPEC200 enclosure as shown in FIGURE 5-8, FIGURE 5-9, and

FIGURE 5-10. TABLE 5-1 shows the color scheme of the SDM wires.

TABLE 5-1. SDM Wiring

Description

Wire Color

EC100

CPEC200

SDM Data

Green

SDM-C1

SDM-C1

SDM Clock

White

SDM-C2

SDM-C2

SDM Enable

Red

SDM-C3

SDM-C3

Digital Ground

Black

Ground

Ground

Shield

Clear

Ground

Ground

To bring cables into the CPEC200 enclosure, remove the cap from

the cable feedthrough by loosening the thumbscrew and twisting

the cap while pulling it off.

To connect a wire to the DIN rail terminal blocks of the CPEC200

enclosure, insert a small screwdriver into the square hole to open

the spring-loaded contacts. Insert the wire into the corresponding

round hole and then remove the screwdriver. Gently tug the wire

to confirm it is secure.

Ensure the CPEC200 enclosure is not powered, and wire the power cable

(CABLEPCBL-L) from the EC100 electronics to the CPEC200 enclosure as

shown in FIGURE 5-8, FIGURE 5-9, and FIGURE 5-10.

Secure the SDM and power cables in the EC100 with a cable tie.

FIGURE 5-8. Wiring of power and communications

NOTE

NOTE

25

Advertising