Appendix C. Output Variables

TABLE C-5. Values stored in table config_history

Variable Name

Units

Description

CO2_SPAN_PPM

µmol∙mol

-1

CO

2

mixing ratio in the CO

2

span cylinder

USE_DIFF_PRESS

Set = True to enable the EC155 differential pressure sensor

CHECK_ZERO

Set = True to enable automatic zero/span sequences

SET_ZERO

Set = True to set the zero during automatic zero/span sequences

CHECK_SPAN1

Set = True to check the CO

2

span during automatic zero/span

sequences

SET_SPAN_1

Set = True to set the CO

2

span during automatic zero/span

sequences

CHECK_SPAN2

Set = True to measure CO

2

span cylinder 2 during automatic

zero/span sequences

CHECK_SPAN3

Set = True to measure CO

2

span cylinder 3 during automatic

zero/span sequences

CHECK_SPAN4

Set = True to measure CO

2

span cylinder 4 during automatic

zero/span sequences

VALVE_MODULE

Compile switch: set = True to enable the use of the valve module

SCRUB_MODULE

Compile switch: set = True to enable the use of the scrub module

SaveAll_diagnostics

Compile switch: set = True to save additional diagnostic values in

the time series data table

Leaf_Wetness_Sensor

Compile switch: set = True to control intake heater voltage based

on a leaf wetness sensor

C-11