Appendix C. Output Variables

If there is some error condition that prevents the zero/span sequence from

starting, one of the following messages will be displayed:

ERROR: Cannot run Zero/Span sequence - CHECK_ZERO is False

ERROR: Cannot run Zero/Span sequence - Check valve temperature

ERROR: Cannot run Zero/Span - Check scrub module temperature

If the CPEC200 begins the startup process for a zero/span sequence, and then

detects an error, one of the following messages will be displayed:

Zero/Span startup manually aborted by turning pump on

ERROR: Zero/Span startup aborted - Check scrub module

temperature

ERROR: Zero/Span startup aborted - Check valve temperature

If the CPEC200 has already started the zero/span sequence and then detects

some error condition that causes the zero/span sequence to abort, one of the

following messages will be displayed:

Zero/Span sequence manually aborted by turning pump on

ERROR: Zero/Span sequence aborted - Check scrub module

temperature

ERROR: Zero/Span sequence aborted - Check valve temperature

If the CPEC200 attempts to set the IRGA zero or span as part of a zero/span

sequence and there is a problem identified (

diag_cpec is nonzero), the

CPEC200 will skip the zero or span setting step, continue the rest of the

sequence, and display one of the following messages. If the user attempts to

set the zero or span manually this error or check is bypassed, and the zero or

span will be set regardless of the status of

diag_cpec.

ERROR: Automatic Zero aborted - Check diagnostics

ERROR: Automatic CO2 Span aborted - Check diagnostics

Some of the messages relate to the use of the valve module in manual mode:

Valve operation not allowed when pump is on

EC mode restarted by turning pump on

ERROR: Valves disabled - Check valve temperature

ERROR: Valves disabled - Check press_offset

ERROR: Valves disabled - Check scrub module temperature

Some of the messages relate to the configuration of the EC155:

ERROR: Powering OFF EC155 gas head

Powering ON EC155 gas head

Configuring the EC100

Setting the zero

Setting the CO2 span

Setting the H2O span

The values stored in table

message_log are shown in TABLE C-4. The value

of the compile switches depends on the constants as described in Section 5.4.2,

C-8