Campbell Scientific CPEC200 Closed-Path Eddy-Covariance System User Manual

Page 62

Advertising
background image

Appendix A. CPEC200 Diagnostics

The purpose of the deadband (the gap between the shutdown voltage and the

turn-on voltage) is to protect the CPEC200 from repeated power cycles when

the battery voltage is very near the shutdown limit.

There are two possible values for

cpec_status when BattVolt_OK = False. If

the battery voltage is below the shutdown limit,

cpec_status will report

ERROR: Battery voltage is too low. In this case, the battery must be recharged

before the CPEC200 will resume normal operation.

If

BattVolt_OK = False and the battery voltage is in the deadband,

cpec_status will report ERROR: Battery voltage is in deadband. This situation

can arise if the battery voltage was previously below the shutdown limit

causing the CPEC200 to set the battery voltage error flag and shut the system

down. Subsequently the battery voltage recovered enough to be in the

deadband (above the shutdown limit but below the turn-on limit).

If

BattVolt_OK = False and BattVolt is in the deadband, manually set

BattVolt_OK = True which may allow the CPEC200 to power up. However,

the increased current drawn upon powerup may cause the battery voltage to fall

below the shutdown limit causing the CPEC200 to shut down again. In this

case, the battery may need to be further recharged.

The voltage at the CPEC200 power input terminals can also be measured and

compared to

BattVolt.

BattVolt is measured in the datalogger which has a diode for

reverse voltage protection. The actual voltage at the input

terminals will be approximately 0.3V higher than the value

reported by

BattVolt.

The CPEC200 supply voltage must be 10.5 Vdc to 16.0 Vdc. See Section 4.3,

Specifications, for details on the power required, especially the extra power

required for a cold startup.

Bit 8: IRGA

If the remainder of

diag_cpec (after subtracting the numeric value for higher

bits that are set) is greater than 127, this means that bit 8 of

diag_cpec is set.

To decode other diagnostic bits, subtract 128 from

diag_cpec and compare the

remainder to the bit values below.

Bit 8 of

diag_cpec is equivalent to irga_OK = False. This indicates either a

diagnostic has been set in the EC100 or that the signal levels are too low:

1. Check the value of

diag_irga. If it is NAN, this indicates the CR3000

is not receiving data from the EC100. Check the SDM cable

connection between the EC100 and the CR3000, and make sure power

is supplied to the EC100.

2. If

diag_irga is −1, this indicates the EC100 does not recognize that an

IRGA is connected. Make sure the EC155 sensor head cable is

connected to the EC100. (Remove power from the EC100 before

connecting or disconnecting the EC155).

NOTE

A-10

Advertising