Campbell Scientific RDP500 Remote Data Platform User Manual

Page 19

Advertising
background image

RDP500 Remote Data Platform

4. Carefully reinstall the catch basin and wait for 5 minutes until the

“Precip_in” field is stable. This delay is to allow post processing in the
datalogger program to sweep from one end of the measurement range to
another.

5. The “Precip_in” field can be viewed by pressing enter on the keyboard

display two times from its default Campbell Scientific start up screen
(Figure 4-2) or by pressing the right arrow key twice on the CD295
external display screen.

FIGURE 4-2. The Precip_in field displayed on a CR1000KD.

6. To initiate the calibration procedure, navigate to the “Initial Setup” menu

in the CR1000KD, where you will find “Step 1”, “Step 2”, and “Cal
Results” (Figure 4-3).

FIGURE 4-3. The Initial Setup menu on the CR1000KD.