Linked test options – Arizona MAX-5000XL User Manual

Page 33

Advertising
background image

AZI Customer Service 800-528-7411 or 602-470-1414

Page 33 of 92

Linked Test Options

Linking tests enables multiple adjoining memory starts to be run together as one test.
Increasing temperatures, rates, times, etc., may be selected to monitor the progress
of moisture loss for research and/or control purposes. In some materials, such as
gypsum powder, moisture occurs in more than one molecular state, usually free water
and one or two bound water

states. A third condition exists in the ashing process

when other components of the sample are driven off at very high temperatures and
only the ash remains.

The Linked Test Option will also prevent rehydration of the sample between tests. As
the test for free moisture is completed, the next memory start in sequence will be
initiated. This subsequent memory start must contain the parameters for the bound
water test. As this test ends, the next memory start may be initiated and so on for as
many links as needed for the process. The sample integrity will not be compromised
between tests since the instrument’s lid is not opened.

Free water evaporates from the sample at relatively low temperatures, typically

40

to 80

C.

Bound water however requires higher temperatures, some around 135

C and

others around 240

C to break the molecular bonds.

The MAX 5000XL also provides temperatures to 600

C for ashing purposes.


The linked tests may also be helpful in determining the temperature at which moisture
is freed in unknown sample material. Setting the temperature a few degrees higher in
each succeeding test will eventually reach a point of weight loss due to evolved
moisture.

Link To Next Test - To run a single

test or to make this the last segment of
linked series, toggle this parameter to
NO.

Result Type - The two selections for

result type offer an Individual
determination for this memory start or
a Total determination of two or more
linked memory starts.
Individual Result

– This selection

provides the result of this test when
two or more tests are linked.

Total Result

– This selection

provides the result using the start
weight of a preceding memory start and the final weight of this memory start.

Advertising