13. MAX 5000XL SPECIFICATIONS

Feature

Specification

Size

18.5” length x 12.7”width x 9.5” height
470 mm length x 323 mm width x 240 mm height

(Height varies slightly with adjustable feet for leveling purposes.)

Weight

31 lbs. (14 kg)

Casework

Textured aluminum castings

Power Requirements 100-130 VAC, 50/60 Hz @ 8 Amps

220-240 VAC, 50/60 Hz @ 4 Amps
Fused on/off switch as part of an EMI power entry module

Operating
Environment

0-35º C @ 50% humidity
0-30º C @ 80% humidity
Non-condensing, non-explosive

Moisture Range

0.1

– 99.9%

Heating Range

25º C to 600º C

Heater Calibration

Menu Driven
NIST Traceable using an accessory Temperature

Calibration Interface module

Calibration tolerance:

± 1.5º C up to 275º C
± 2º C up to 600º C

Twelve calibration points from 50º C to 575º C.
Twelve preset verification points from 50º C to 575º C, or

up to four user-programmable verification points.

Programmable calibration reminder

Temperature Control

± 1º C at temperatures between 50º C and 300º C
± 2º

C at temperatures between 301º C and 600º C

Test Temperatures

User programmable through front panel keypad to control

idle temperature, test temperature, and high start
temperature for all tests.

Multiple test temperatures are programmable for linked

test sequences

Heater Chamber

Composite ceramic fiber lined heater chamber
Isolated from electronics to minimize damage from

corrosive sample materials

Oven self cleaning cycle (550º C for 45 minutes)

Balance

Integrated balance with automatic operation
Balance capacity to 100 grams
Resistant to nominal vibration, with options available for
isolation feet
Resistant to air currents from internal exhaust fan
Isolated from extreme temperatures in the test chamber
Isolated from extreme changes in humidity caused by
evaporation in the test chamber