Standard Commands for Programmable Instruments (SCPI)

100

Query Syntax

STAT

us

:OPER

tion

:EVEN

t?

Returned Parameters

<NR1>

(Register Value

)

Examples

STAT: OPER: EVEN? STATUS: OPERATIONAL: EVENT?

NTR/PTR Commands

These commands allow you to set or read the value of the Operation NTR (Negative-
Transition) and PTR (Positive-Transistion) registers. These registers serve as polarity filters
between the Operation Enable and Operation Event registers to cause the following actions:

When a bit of the Operation NTR register is set to 1 then a 1-to-0 transition of a corresponding enabled bit of Operation
Condition register will cause that bit to be set in the Operation Event register.

When a bit of the Operation PTR register is set to 1 then a 0-to-1 transition of a corresponding enabled bit of Operation
Condition register will cause that bit to be set in the Operation Event register.

If the same bits in both NTR and PTR registers are set to 1, then

any transition

of that bit (if enabled) at the Operation

Condition register will set the corresponding bit in the Operation Event register.

If the same bits in both NTR and PTR registers are set to 0, then

no transition

of that bit (even if enabled) at the

Operation Condition register can set the corresponding bit in the Operation Event register.

NOTE

I

f a Operation bit is in the proper state, then programming the corresponding1 PTR or NTR filter bit to

1

will set the associated Event Register bit.

Command Syntax

STATus:

OPER

tion

:NTR

ansition

<NRf>

STAT

us

:OPER

tion

:PTR

ansition

<NRf>

Parameters

0 to 32727

Default Value

0

Examples

STAT:OPER:NTR32 STATUS:OPERATION:PTR 1312

Query Syntax

STAT:OPER:NTR? STAT:OPER:PTR?

Returned Parameters

<NR1>

(Register value

)

STAT:PRES

This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsytem NTR and
Enable registers. STAT:OPER:PTR is set to 1313 and STAT:QUES:PTR is set to 1555.

Command Syntax

STAT

us

:PRES

et

Examples

STAT: PRES

STATUS: PRESET

Status Questionable Registers

The bit configuration of all Status Questionable registers is as follows:

Bit Configuration of Questionable Registers

Bit Position

15-11

10

9

8

7

6

5

4

3

2

1

0

Condition

NU

UNR

RI

NU

NU

NU

NU

TO

NU

NU

OC

OV

Bit Weight

1024

512

256

128

64

32

16

8

4

2

1

NU = (Not used); OC = Overcurrent protection circuit has tripped; OT = Overtemperature status condition exists;
OV = Overvoltage protection circuit has tripped; RI = Remote inhibit is active; UNR = Power supply output is
unregulated.

See "Status Reporting" for more explanation of these registers.