Agilent Technologies 6028A User Manual

Page 91

Advertising
background image

Standard Commands for Programmable Instruments (SCPI)

91

Description

This command causes the interface to set the OPC bit (bit 0) of the Standard Event Status register when the power supply
has completed all pending operations. (See

*ESE

for the bit configuration of the Standard Event Status register.)

Pending operations

are complete when:

all previous commands have been executed

any change in the output level caused by previous commands has been completed (completion of settling time, relay
bounce, etc.)

no pending trigger level operations are set

*OPC

does not prevent processing of subsequent commands but Bit 0 will not be set until

all pending operations are

completed.

Command Syntax

*OPC

*OPC?

Meaning and Type

Operation Complete

Device Status

Description

This query causes the interface to place an ASCII "1" in the Output Queue when all pending operations are completed.

Pending operations

are as defined for

*OPC

. Unlike

*OPC, *OPC?

prevents processing of all subsequent commands.

*OPC?

is intended to be used at the end of a command line so that the application program can then monitor the bus for

data until it receives the "1" from the power supply Output Queue.

Do not follow

*OPC?

with

*TRG

or GP-IB bus triggers. Such triggers sent after

*OPC?

will be prevented

from executing and will stop system operation. If this occurs, the only programmable way to restore
operation is by sending the power supply a GP-IB DC1 (Device Clear) command.

Query Syntax

*OPC?

Returned Parameters

<NR1>

ASCII 1 is placed in the Output Queue when the

power supply has completed operations.

*PSC

Meaning and Type

Power-on Status Clear

Device Initialization

Description

This command controls the automatic clearing at power turn on of:

the Service Request Enable register

the Standard Event Status Enable register