Standard Commands for Programmable Instruments (SCPI)

98

OUTP:PROT

There are two output protection commands that do the following:

OUTP:PROT:CLE

Clears any OV (overvoltage), OC (overcurrent, unless set via external voltage control),OT

(overtemperature), or RI (remote inhibit) status conditions. If the power supply is configured with
the Agilent 59510A/llA Relay Accessory, it will be restored to the state it was in before the
protection circuit tripped.

OUTP:PROT:DEL

Sets the time that a change in CV, CC, or UNREG condition must appear at the output before that

change is recorded in the Status Operation Condition register. The delay prevents momentary output
changes that can occur during reprogramming from being registered as events by the status
subsystem.

Command Syntax

OUTP

ut:

PROT

ection:

CLE

ar

OUTP

ut:

PROT

ection:

DEL

lay <NRf+>

Parameters

OUTP:PROT:CLE

(None)

OUTP:PROT:DEL

See "DI,Y" in Table 3-7 of Section III.

Default Sufflx

S

Examples

OUTP: PROT: CLE OUTPUT: PROTECTION: CLEAR
OUTPUT:PROTECTION:DELAY 75E-l
OUTP:PROT:DELMIN OUTPUT:PROT:DELAY MAX

Query Syntax

OUTP:PROT:CIE

(None

) OUTP:PROT:DEL?

OUTP:PROT:DEL? MIN OUTP:PROT:DEL? MAX

Returned Parameters

<NR3> OUTP:PROT:DEL

returns value of programmed delay

.

OUTP:PROT:DEL? MIN

and

OUTP:PROT:DEL? MAX

return the minimum and maximum

programmable delays.

OUTP:REL[:STAT]

This command is valid only if the power supply is configured for the optional Agilent 59510A and Agilent 59511A Relay
Accessory. Programming

ON

closes the relay; programming

OFF

opens it. If the power supply is not configured for either

device, the command generates an error.

Command Syntax

OUTP

ut

:REL

ay

[:STATe] <bool>

Parameters

0 or OFF

I or ON

*RST Value

0

Examples

OUTP: REL 1 OUTPUT: RELAY OFF OUTPUT: RELAY: STATE OFF

Query Syntax

OUTP:REL?

Returned Parameters

<NR.1> 0 or 1

OUTP:REL:POL

This command is valid only if the power supply is configured for the Agilent 59511A Relay Accessory. The

NORM

al

command causes the device output polarity to agree with that of the power supply. Programming

REV

causes the polarity of

the device to be opposite to that of the power supply. If the power supply is not configured for the Agilent 59511A, either
command generates an error.

Command Syntax

OUTP

ut

:REL

ay

:POL

arity

{NORMalREV

erse

}

Parameters <CRD>

NORMal

or

REVerse

*RST Value

NORM

Examples

OUTP: REL: POL NORM OUTPUT: RELAY: POLARITY REVERSE

Query Syntax

OUTP:REL:POL?

Returned Parameters

<CRD> NORM

or

REV