Standard Commands for Programmable Instruments (SCPI)

99

Status Subsystem

This subsystem programs the power supply status registers. The power supply has two groups of status registers;

Operation

and

Questionable.

Each group consists of the following five registers:

Condition

Enable Event

NTR Filter

PTR Filter

Status Operation Registers

The bit configuration of all Status Operation registers is shown in the following table:

Bit Configuration of Operation Registers

Bit Position

15-12

11

10

9

8

7

6

5

4

3

2

1

0

Condition

NU

NU

CC

NU

CV

NU

NU

WTG

NU

NU

NU

NU

NU

Bit Weight

2048

1024

512

256

128

64

32

16

8

4

2

1

CC = Power supply is sourcing current in CC mode; CV = Power supply is in constant voltage mode; NU = (Not used);
WTG = Interface is waiting for a trigger.

See "Status Reporting" for more explanation of these registers.

STAT:OPER:COND?

Returns the value of the Operation Condition register. That is a read-only register which holds the real-time (unlatched)
operational status of the power supply.

Query Syntax

STAT

us

:OPER

ation

:COND

ition

?

Examples

STAT: OPER: COND? STATUS: OPERATION: CONDITION?

Returned Parameters

<NRl>

(Register value)

STAT:OPER:ENAB

Sets or reads the value of the Operation Enable register. The Operation Enable register is a mask for enabling specific bits
in the Operation Event register that can cause the operation summary bit of the Status Byte register to be set. The operation
summary bit (bit 7) is the logical OR of all the enabled bits in the Operation Event register.

Command Syntax

STAT

us

:OPER

ation

:ENABl

e

<NRf>

Parameters

0 to 32727

Default Value

0

Examples

STAT: OPER: ENAB 1312 STAT: OPER: ENAB 1 STATUS:OPERATION:ENABLE? Query

Syntax

STAT:OPER:ENAB?

Returned Parameters

<NRl>

(Register value)

STAT:OPER:EVEN?

Returns the value of the Operation Event register. The Event register is a read-only register
which holds (latches) all events that:

are enabled by the Operation Enable register.

are passed by the Operation NTR and/or PTR filter.

Reading the Operation Event register clears it.