Standard Commands for Programmable Instruments (SCPI)

93

*RST

Meaning and Type

Reset

Device State

Description

This command reset the power supply to a factory-defined state as defined below.

*RST

also

forces an

ABORt

and an

OUT:PROT:CLE

command.

CURR[:LEV] [:IMM] *

OUTP OFF

VOLT[:LEV][:IMM]*

CURR[:LEV]:TRIG*

OUTP:PROT:DEL*

VOLT[:LEV][:TRIG]*

CURR:PROT:STAT OFF

OUTP:REL OFF

INIT:CONT OFF

OUTP:REL:POL NORM

*Model-dependent value. See

ISET, DI,Y,

and

VSET

in Table 3-7 of Section III..

Command Syntax

*RST

*SAV

Meaning and Type

SAVE

Device State

Description

This command stores the present state of the power supply to the specified location in memory. Up to five states can be
stored. Because the memory locations are volatile, all saved states are lost when the power supply is turned off. When the
power supply is turned on, all memory locations are initialized to the factory defined state (see the

*RST

command).

The following power supply parameters are stored by

*SAV

:

CURR[:LEV][:IMM]

OUTP[:STAT]

OUTP:REL:POL

CURR:PROT:STAT

OUTP:PROT:DEL

VOLT[:LEV][:IMM]

OUTP:REL[:STAT]

*SAV causes a write cycle to nonvolatile memory. Non volatile memory has a finite maximum number of
write cycles. Programs that repeatedly cause write cycles to nonvolatile memory can eventually exceed the
maximum number of write cycles and cause the memory to fail.

Command Syntax*SAV <NRf>
Parameters

0 to 4

Example

SAV 3