Agilent Technologies 6028A User Manual

Page 76

Advertising
background image

100 VAC Input Power Option 100

76

Section ll Manual Changes

For Agilent Models 6030A, 6031A, 6032A and 6035A: on page 2-7 (AC line Impedance Check, step b), where the
maximum output voltages are tabulated, change the voltages as shown below:

6030A change 65 V to 50 V
6031A change 8 V to 6 V
6032A change 22 V to 13.5 V
6035A change 220 V to 150 V

Section lll Manual Changes

On page 3-10 (Overvoltage Protection), change 107% to 90%.
On page 3-11 (Monitor Signals), change “0-5 Volts” to ‘’0-4.5 Volts’ ‘ .
In Table 3-6, change the soft voltage limits as follows:

Model

from:

to:

6030A

204.75

174.03

6031A

20.475

17.40

6032A

61.425

52.21

6035A

511.875

435.10

6033A

20.475

17.40

6038A

61.425

52.21

In table 3-7, GP-IB Commands, change the boxes in the second column (*Range **Response) with voltages tabulated in
tables a, b, c, or d as shown below.

Change the column for VSETx, VSETxV, VMAXx, and VMAXxV to set a.
Change the column for VSETxmV, and VMAXxmV to set b.
Change the column for ISETxA, ISETxA, IMAXx, and IMAXxA to set c.

Change the column for ISETxmA, and IMAXxmA to set d.

Agilent Model

Set a

Set b

Set c

Set d

6030A

0-175.00 V

0-175000 mV

0-17.403 A

0-17403 mA

6031A

020.45 V

0-20425 mV

0-122.85 A

0-122850 mA

6032A

0-51.495 V

0-51495 mV

0-51.1875 A

0-51187.5 mA

6033A

0-17.500 V

0-17500 mV

0-30.7125 A

0-30712.5 mA

6035A

0-437.50 V

0-437500 mV

0-5.119 A

0-5119 mA

6038A

0-51.495 V

0-51495 mV

0-10.2375 A

0-10237.5 mA