Agilent Technologies 6028A User Manual

Page 50

Advertising
background image

Operating Instructions

50

Table 3-7 GP-IB Commands

Command

*Range or **Response
To Query

Description

Page

VSET x
VSET xV

Agilent 6030A *0--204.75 V
Agilent 6031A *0--20.475 V
Agilent 6032A *0--61.425 V
Agilent 6033A *0--20.475 V
Agilent 6035A *0--511.88 V
Agilent 6038A *0--61.425 V

Any of these commands is used to program

output voltage. Initial Condition: 0 V

55

VSET xMV

Agilent 6030A *0--204750 mV
Agilent 6031A *0--20475 mV
Agilent 6032A *0--61425 mV
Agilent 6033A *0--20475 mV
Agilent 6035A *0--511880 mV
Agilent 6038A *0--61425 mV

ISET x
ISET xA

Agilent 6030A *0--17.403 A
Agilent 6031A *0--122.85 A
Agilent 6032A *0--51.1875 A
Agilent 6033A *0--30.7125 A
Agilent 6035A *0--5.119 A
Agilent 6038A *0--10.2375 A

Any of these commands is used to program

output current. Initial Condition: 0 A

56

ISEFT xMA

Agilent 6030A *0--17403 mA
Agilent 6031A *0--122850 mA
Agilent 6032A *0--51187.5 mA
Agilent 6033A *0--30712.5 mA
Agilent 6035A *0--5119 mA
Agilent 6038A *0--10237.5 mA

VSET?
ISET?

**VSET xx.xxx
**ISET xx.xxx

Used to read voltage and current settings

56

VOUT?
IOUT?

**VOUT xx.xxx
**IOUT xx.xxx

Used to measure and read output voltage
or current.

56

OVP?

**OVP xx.xx

Causes power supply to measure OVP
setting (which is hardware set at front panel).

57

VMAX x
VMAX xV

VMAX xMV

(Same as VSET)

Any of these commands is used to program an
upper limit (soft limit) on the voltage
programming value that the power supply will
accept.

57

IMAX x
IMAX xA

IMAX xMA

(Same as ISET)

Any of these commands is used to program an
upper limit (soft limit) on the current
programming value that the power supply will
accept.

57

VMAX?
IMAX?

**VMAX xx.xxx
**IMAX xx.xxx

Used to read voltage and current limits (soft
limits).

57