Syst:err, Syst:lang – Agilent Technologies 6028A User Manual

Page 102

Advertising
background image

Standard Commands for Programmable Instruments (SCPI)

102

Command Syntax

STAT

us

:QUES

tionable

:NTR

ansition

<NRf>

STAT

us

:QUES

tionable

:PTR

ansition

<NRf>

Parameters

0 to 32727

Default Value

0

Examples

STAT:QUES:NTR 16 STATUS:QUESTIONABLE:PTR 512

Query Syntax

STAT:QUES:NTR? STAT:QUES:PTR?

Returned Parameters

<NR1>

(Register value)

SYST:ERR?

Returns the next error number followed by its corresponding error message string from the remote programming error
queue. The queue is a FIFO (first-in, first-out) buffer that stores errors as they occur. As it is read, each error is removed
from the queue. When all errors have been read, the query returns 0,N0 ERROR. If more errors are accumulated than the
queue can hold, the last error in the queue will be -350,T00 MANY ERRORS.

Query Syntax

SYST

em

:ERR

or

?

Returned Parameters

<NRl>,<SRD>

Examples

SYST: ERR? SYSTEM: ERROR?

SYST:LANG

Switches the interface between its SCPI (TMSL) command language and it compatibility language. Note that the earlier
parameter "TMSL" must be used in place of "SCPI". Sending the command causes:

The alternate language to become active and to be stored in nonvolatile memory.

The power supply to reset and clear the storage and error registers.

NOTE

After sending the SYST:LANG command, do not send another command or turn the supply off for at least
five seconds

.

If the power supply is shut off, it will resume operation in the last-selected language when power is restored.

Command Syntax

SYST

em

:LANG

uage

{TMSL|COMP

atibility

}

Syntax is the same, regardless of the present language

Parameters

<CRD> TMSL

or

COMPatibility

Default Value

TMSL or last selected language

Examples

SYST:LANG TMSL SYSTEM:LANGUAGE COMPATIBILITY

Query Syntax

SYST

em

:LANG

uage

?

Returned Paramters

<CRD> TMSL

or

COMP

Advertising
This manual is related to the following products: