APPENDIX 1 - SOFTWARE LOGCHART-II

INSTALLING LOGCHART-II
LogChart II is the software provided with the logger to allow for

configuration and data offload. To install, run the LC_II_Setup.exe

file provided with the CD.
Note: Be sure your Windows date separator is configured as a slash:

dd/mm/yy or dd/mm/yyyy.

RUNNING LOGCHART-II
When you open LogChart-II the main window is displayed:

Fig. 1 – LogChart-II main window

Then indicate the serial port to be used by the communication

interface through the "Port" menu. This menu lists all available serial

port on your computer. You must select a port that shows the N120

equipment in their description.

Fig. 2 – Selection of USB serial port where the N120 controller is connected.

Once a valid serial port is selected, the icons for communication with the

controller will be activated.

Fig. 3 – Icons enabled when the communication port selected is a valid port

CONFIGURING THE CONTROLLER

To configure the controller press the button:

The Configuration Settings window is displayed. This LogChart-II

window allows the user to set the operating mode of the controller

and also provide general information about the device (Fig. 4).

Fig. 4 – Logger Set Up (Readings)

The configuration fields are:

Title: In this field, the user identifies the records giving it a

name (up to 16 characters).

General Information: Information field. There are presented

information regarding the controller, including: Model (N120), serial

number, date/time of the controller, current date/time of the PC,

firmware version (model controller), memory capacity and number of

acquisitions in memory.
In this field the time is constantly updated while the communication

between controller and computer is established.
Note: The clock of the controller may be with a little difference

compared to the current time of the computer. When the controller is

configured their clock is updated.

Acquisitions: Presents a series of parameters that define how

measurements will take place.

Interval: Defines the interval between readings: Minimum

interval is 1 (one) second.

Estimated time: In this parameter, the logger informs the user

how long it will take to occupy the full memory, in the conditions

set up during configuration.

Start of Readings: Readings can be started in one of five

different modes:

Immediately: start as soon as programming is considered

to be ready, and is then sent (OK) to the logger.

Digital Input: The acquisitions are started whenever the

digital input of the N120 is activated. Otherwise, the

acquisitions are interrupted

Alarm: the acquisition starts whenever the condition

associated with the alarm 1 (AL1) controller is met.

Otherwise, stops the acquisition.

End of the acquisitions: The options for the ending of

acquisitions are:

Memory Full: acquisitions are made up to the capacity of

the memory available.

No Stop (Circular Memory): acquisitions happen continuously,

overwriting the oldest registers as the number of acquisitions

exceeds the memory capacity.

Channels: Shows parameters referring to each channel

separately. Channel 1 is the Temperature channel, and Channel 2

the Relative Humidity channel.

Fig. 5 – Logger Set Up (Channels)

Tag: Defines a name for the temperature registers.

Offset: Makes possible to correct the value logged.

Valor: It defines how the value measured will be registered.

Unit: Defines the unit of the value measured: ºC or ºF for

channel 1 (temperature).

Entrada: The alarm limit value is shown with a dotted line with

the same color of the channel 1.