5.3. PROTECTION
The MODEL DC80L calibrator uses two protection fuses, rated 125 mA / 250 V,

internally connected in series with the current terminals (2) and (4).
To avoid personal injury or damage to the calibrator, use only 125 mA / 250 V fast

fuses.

Notes on fuses:

- Fuse 1 is probably blown out when the display flashes the message OL while

in Voltage output mode with the leads removed from the calibrator.

- Fuse 2 is probably blown out if the calibrator reads 0.000 in mA input mode,

even under the presence of a current input signal

5.4. MODEL DC80L OPERATING INSTRUCTIONS
DC mV and V measurements

1. Press the power switch to

turn on the Calibrator.

2. Press the INPUT / OUTPUT

mode key to select INPUT

mode.

3. Press the V/mV mode key

to select mV input .

4. Plug the red test lead to “V” jack

and the black one to the

COM” jack .

5. Connect the test leads

according to the signal polarity:

red to the positive polarity, black

to the negative polarity (GND).

6. Measurement value is shown on

the display .