Dc ma generation – NOVUS DC80L User Manual

Page 12

Advertising
background image

12

b) Measurements using calibrator’s internal power supply for powering the

loop

The calibrator provides an internal 24 V power supply for measuring the signal of

2-wire transmitters (sink transmitters). The 24 Vdc is applied between terminals

+LOOP e COM. The sink transmitter is to be connected between terminals

+LOOP (Power supply output) and mA (terminal for the loop current monitoring).

1. Press the power switch

to

turn on the Calibrator.

2. Choose OUTPUT mode by

pressing the INPUT / OUTPUT

key .

3. Press the mA / mA % mode key

to choose the desired unit.

The sings mA or mA %

will

be presented on the display.

4. Plug the red test lead into the

LOOP jack

and the black

one to the mA jack .

5. Connect the test leads such that

the current enters the red lead

and returns to the power supply

through the black lead.

6. The measurement is shown on

the display .


DC mA generation
a) Sourcing mA

1. Press the power switch to

turn the Calibrator ON.

2. Press the input / output mode

key

to select OUTPUT

mode .

3. Press the mA / mA % mode key

to select mA or mA % .

4. Press the increment

/

decrement keys

to set the desired output value.

5. Plug the red test lead into the

LOOP jack

and the black

one to the V jack .

6. Connect the test leads such that

the current enters the red lead

and returns to the power supply

through the black lead.

7. To change the output value,

press the increment / decrement

keys

and / or

the mA mA % mode key .


Advertising
This manual is related to the following products: