Model dc80t operating instructions, Thermocouple or mv measurement, Thermocouple or mv simulation – NOVUS DC80L User Manual

Page 4

Advertising
background image

4

3.3. MODEL DC80T OPERATING INSTRUCTIONS
In thermocouple temperature simulation the calibrator generates the voltage as an

actual thermocouple, considering both the simulated temperature (hot joint) and

connection temperature (cold joint). To ensure stated accuracy the user must use

a connection cable between calibrator and calibrated device with the same

thermoelectric properties of the selected thermocouple (extension or

compensation cable).
A K type extension cable (yellow connector) is supplied with the calibrator and

should only be used for K type calibration for maximum accuracy.
Make sure that the appropriate cable with the correct polarity is being used to

avoid large calibration errors.

Thermocouple or mV measurement

1. Press the power switch to turn

on the calibrator.

2. Press the INPUT / OUTPUT key

to select INPUT mode.

3. Press the TC TYPE selection key

to configure the instrument to the

desired input type

4. Connect the T/C or mV signal to the

input connector .

5. The measured value is presented in

.

Thermocouple or mV Simulation

1. Press the power switch to

turn on the calibrator.

2. Press the INPUT / OUTPUT key

to select OUTPUT mode.

3. Press the TC TYPE selection

key

to configure the desired

output signal type.

4. Press the increment /

decrement keys

to adjust the value to be

simulated.

5. The simulated signal is

delivered at the output

connector .

6. To change output values, press

the increment / decrement keys

until the desired value is

presented in the display.

Changing to other input types is

accomplished by the TC TYPE

selection key .

Advertising