SMARTMETER

PORTABLE TEMPERATURE METER - INSTRUCTIONS MANUAL – V1.0x

1 INTRODUCTION

SmartMeter is a portable temperature meter with high-contrast LCD.

The Smart Meter housing enables industrial environment or bench-

top operation and provides a belt carrying fixture. It supports 2

temperature sensors with independent and concurrent display of

measured values. It provides the capability of reading 0-50 mV

electric signals with configurable indication scale.
Difference between channels, relative measure, minimum, maximum

and average values, hold and alarm functions are also available. It

also features an auto power off resource in case the device is idle, in

order to extend battery life.
The device is supplied with a K-type thermocouple with 1.5 m cable.

1.1 UNPACKING
The meter is supplied in proper packaging that offers suitable

protection during transit and storage. Unpack with care. Upon

delivery, make sure that there is no damage to the instrument, and

that there are no missing items.
Within packaging, you should find, besides the manual, the meter and

a K-type thermocouple sensor.

2 SPECIFICATIONS

Input types and max. indication range

J thermocouple ..........................................................

- 100 to 760 ºC

K thermocouple.......................................................

- 150 to 1.370 ºC

T thermocouple..........................................................

- 160 to 400 ºC

E thermocouple.............................................................-90 to 720 ºC
N thermocouple ......................................................

- 270 to 1,300 ºC

R thermocouple ........................................................

- 50 to 1,760 ºC

S thermocouple.........................................................

- 50 to 1,760 ºC

B thermocouple.........................................................500 to 1,820 ºC
Pt100 .........................................................................

- 200 to 600 ºC

Voltage 0-50 mV...................................

- 1,999 to 9,999 (Adjustable)

All input types are factory calibrated. NBR 12771, Pt100 NBR

13773/97 (IEC-751) (385) calibrated thermocouples.
Accuracy:

Thermocouples: ..............................0.25 % F.S.

± 1 ºC

Pt100:................................................0.2 % F.S.

± 1 ºC

0-50 mV: .........................................0.2 % F.S.

± 1

digit

Resolution:......................................................................................0.1 ºC
Sampling rate:............................................................. 2 timer per second
Input impedance: ...............................> 10 M (Thermocouples, 50 mV)
Pt100 excitation current:............................................................0.165 mA
Power supply:......................................................................9 Vdc battery
AC / DC adapter .............................................................................6 Vdc
Estimated autonomy:.................................... 400 hours (alkaline battery)
Dimensions:................................................................. 165 x 73 x 36 mm
Weight:.............................................................................................205 g

Protection

: ..........................................................................................IP40

3 CONNECTIONS

Female input connectors suitable for RTD / TC compensated

connectors, as shown below.

Figure 01 — Meter supported connectors

Only two input terminals are used to connect thermocouples. Input

terminals have different size holes, which prevent inappropriate

installations.

Figure 02 — Thermocouple connection

All three terminals are used for Pt100 (RTD)

Figure 03 – Pt100 connectors sensors

The Pt100 internal connections should be made as shown in figure

below.

Figure 04 — Internal connections of Pt100 sensors

For the 0-50 mV signal, installation should comply with the following

connections:

Figure 05 — 0-50 signal connection