4 operation, Alarm cycle – NOVUS SMARTMETER User Manual

Page 2

Advertising
background image

SmartMeter

NOVUS PRODUTOS ELETRÔNICOS LTDA.

2/4

4 OPERATION

The meter features two input channels. Channel readouts are

updated two times per second. Channel 1 readouts are shown in the

largest digits and channel 2 readouts are shown in the smallest digits,

as Figure 07 shows. Channel 02 can be turned-off by operator.

Figure 06 — Meter front panel

Figure 07 — SmartMeter display

4.1 TURNING THE METER ON/OFF
Press

to turn on the meter. To turn off, press for 3 seconds.

The meter basic configuration is factory set, so it can be immediately

used with the sensor provided. Factory configuration is: two enabled

channels, K-type sensor, one decimal space, zero offset, sound

alarm not enabled, auto power off after five minutes, temperature unit

in degrees Celsius, 60 Hz filter and channels 1 and 2 for special

functions.

4.2 FIRST READOUTS
Insert the sensor provided with the meter in the terminals of the

required channel and take the sensor leads to the measurement

location. Check the indication displayed.
Use the keys to find special functions the meter offers. When using

other sensors, check if factory setup is still compliant. If not, access

the configuration cycle and change settings.

4.3 SPECIAL FUNCTIONS
HOLD FUNCTION
It “freezes” the measure reading indication. HOLD is then displayed.
It is activated by pressing key

HOLD

.

To resume the meter operation,

press key

HOLD

again.

MIN FUNCTION
It allows the user to retrieve the

minimum temperature value

measured within an interval. An indicator of such condition is
displayed in the meter. It is activated by pressing key

AVG

MAX MIN

.

To

resume the meter operation, press key

AVG

MAX MIN

again.

MAX FUNCTION
It allows the user to retrieve the

maximum temperature value

measured within an interval. An indicator of such condition is
displayed in the meter. It is activated by pressing key

AVG

MAX MIN

.

To

resume the meter operation, press key

AVG

MAX MIN

again.


AVG FUNCTION (average)
It displays the

average temperature value measured within an

interval. An indicator of such condition is displayed in the meter. It is
activated by pressing key

AVG

MAX MIN

.

To resume the meter operation,

press key

AVG

MAX MIN

again.

Notes: Functions MIN, MAX and AVG are accessed via key

AVG

MAX MIN

.

Values displayed are the ones acquired from the start of an interval of

measurements. The interval starts after a RESET action. To reset the

meter, press and hold key

AVG

MAX MIN

for more than 10 seconds. When

reset button is pressed, MIN, MAX and AVG values are erased and a

new interval starts.
RELAT FUNCTION (relative)
When enabled, it will void the measured values and starts to indicate

variations from such zeroed values. An indicator of such condition is
displayed in the meter. It is activated by pressing key

RELAT

.

To

resume the meter operation, press key

RELAT

again.

T1 – T2 FUNCTION
When enabled, it displays the highest value corresponding to the

difference between the value measured through channel 1 and the

value measured through channel 2. The indication corresponding to

channel 2 is not changed. An indicator of such condition is displayed
in the meter. It is activated by pressing key

T1 – T2

.

To resume the

meter operation, press key

T1 – T2

again.

Note: This function is only enabled when both channels are enabled

and operating.
ALARMS
The meter has an audible alarm (BEEP) that can inform the operator

about an alarm condition. In each channel there can be two indication

values, when they are reached the audible alarm is triggered. A visual

indicator is also displayed in the meter when the indication is under

an alarm condition.
LOW BATTERY INDICATOR
A low meter indicator is displayed warning that the battery should be

replaced.
SET UP
To set up the meter, the user must access the Alarm and Meter Set

up cycles, where a series of parameters must be defined.

ALARM CYCLE

In the alarm cycle you find the setpoints parameters of the meter

alarms. Press

SET

to access this cycle. The meter displays

,

the first parameter of the cycle. To access the remaining parameters
of this cycle, press

SET

again

,

until you reach the measures screen

.

Keys

and

make possible to change the parameters defined.

The parameters of this cycle are:

(Highest setpoint of channel 1)

Determines a temperature value in channel 1 that when exceeded

upwards triggers the meter alarm (BEEP). When the alarm is

triggered, the indication

ALARM

is displayed in the meter.

(Lowest setpoint of channel 1)

Determines a temperature value in channel 1 that when exceeded

downwards triggers the meter alarm (BEEP). When the alarm is

triggered, the indication

ALARM

is displayed in the meter.

(Highest setpoint of channel 2)

Determines a temperature value in channel 2 that when exceeded

upwards triggers the meter alarm (BEEP). When the alarm is

triggered, the indication

ALARM

is displayed in the meter.

(Lowest setpoint of channel 2)

Determines a temperature value in channel 2 that when exceeded

downwards triggers the meter alarm (BEEP). When the alarm is

triggered, the indication

ALARM

is displayed in the meter.